ANAESTHETIC RESERVOIR BAGS
The anaesthetic reservoir bags produced by the companystrictly comply with the YY/T0978-2016/ISO5362:2006
standard. The anaesthetic reservoir bags are classified into four categories according to the raw material: natural
rubber, neoprene, synthetic polyisoprene rubber and nitrile rubber. According to the capacity, the products are divided
into 0.5 liters, 1 liter, 2 liters, 3 liters, and multiple specifica- tions. The company can also customize special products
according to customer needs.
PRODUCT DETAILS
Product performance | |
appearance | The breathing bag used in anesthesia machines should have consistent color and be free of defects such as spots, stains, pockmarks, and burrs. |
capacity | The actual capacity error of the breathing bag used in anesthesia machines should be within ± 15% of the nominal value. |
leakage | a. The nominal capacity of the breathing bag used in anesthesia machines shall not exceed 1L, The leakage under internal pressure of (3 ± 0.3) kPa should not exceed 10mL/min. b. If the nominal capacity of the breathing bag used in anesthesia machines exceeds 1L, The leakage under internal pressure of (3 ± 0.3) kPa should not exceed 25mL/min. |
Airbag connector | a. The neck of the breathing bag used in anesthesia machines is a naked neck structure. b. The inner cone joint meets the requirements of the 22mm specification cone joint in YY1040.1-2003. |
Capsular tail | The tail length of the breathing bag used in anesthesia machines is greater than 20mm. |
Capsule anti dilation (Pressure/Volume) Required pressure | a. The final pressure shall not be less than 3.0kPa and not more than 6.0kPa. b. It should be restored to its initial measured volume within 30 minutes of the experiment, The error is within ± 10% of the initial measured volume. |
Product color
Product specifications
HH-HHN-A-0.5L
HH-HHN-A-1L
HH-HHN-A-2L
HH-HHN-A-3L
HH-HHN-B-0.5L
HH-HHN-B-1L
HH-HHN-B-2L
HH-HHN-B-3L